Download Section 1: Introduction

Transcript
QUADRA Q4 Diamond Series II
7.2 QUADRA Q4 DIAMOND SERIES II Error Messages
Error Message
Potential Cause
INTERLOCK OPEN - REPLACE
SYSTEM DISCHARGED
Remote Interlock on rear of machine
disconnected or dummy plug missing.
Connect remote interlock or replace
dummy plug.
HANDPIECE OPEN - CLOSE
SYSTEM DISCHARGED
Front of handpiece open or incorrectly
closed.
Ensure disposable unit located
correctly and close handpiece.
LAMP nn FAILURE
PRESS ANY KEY TO CLEAR
Flashlamp misfire.
Clear fault and retry. If problem
persists, replace disposable unit and
retry.
Incorrect or unapproved disposable
inserted into handpiece.
Replace disposable unit and retry.
Failure to communicate with
disposable.
Remove disposable and refit. If
problem persists, replace disposable
unit and retry.
BAD DISPOSABLE DETECTED
SWITCH OFF AND REPLACE
FAULTY DISPOSABLE
SWITCH OFF AND REPLACE
Remedy / Action
FAILED TO CHARGE SYSTEM
SWITCH SYSTEM OFF
DISCHARGE RELAY CLOSED
SWITCH SYSTEM OFF
INCORRECT CAP VOLTS
SWITCH SYSTEM OFF
HEAD OVER TEMPERATURE
SWITCH SYSTEM OFF
SYSTEM OVER TEMPERATURE
SWITCH SYSTEM OFF
Switch off Quadra Q4 Diamond Series II Unit. Wait for 2 minutes. Switch on the
Quadra Q4 Diamond Series II following standard start-up procedure.
If problem persists, call Service Operator.
SYSTEM 30V SUPPLY FAILED
SWITCH SYSTEM OFF
HEAD COOLING FANS FAILED
SWITCH SYSTEM OFF
DISCHARGE RELAY OPEN
SWITCH SYSTEM OFF
TOP LAMP MISFIRE
Flashlamp misfire
BOTTOM LAMP MISFIRE
Page 40
- Check Pins.
- Replace disposable unit and retry.