Download Manual
Transcript
• Only use spare parts supplied by the device manufacturer. • Make sure your mains supply specifications meet the electrical requirements for the device indicated on the label and in this manual. • Do not use several devices at once in addition to the SkinBase IPL system. • Do not use the device in environments with high electromagnetic fields as these could cause the SkinBase IPL or other nearby devices to malfunction. • Do not use mobile phones in the treatment area. • Carry out device maintenance according to the manufacturer’s instructions. • The device must be installed and put into service according to the instructions contained in this manual • Portable and mobile radio communication devices may affect the operation of the device • Do not operate in proximity of a short-wave or microwave therapy APPARATUS (e.g. within 1m); • Do not connect the patient to a high-frequency electrosurgery apparatus simultaneously; SPECIFICATIONS SkinBase IPL is a high intensity pulsed light system intended for hair removal and photorejunevation. This treatment is based on the scientific principle of selective photothermolysis, which treats only the selected area without damaging the surrounding tissue. SKINBASE IPL IS DESIGNED FOR CONTINUOUS TREATMENT OVER SHORT INTERMITTENT PERIODS ACCORDING TO THE FOLLOWING TIMES: T-ON=4 MIN AND T-OFF=5 MIN. DURING TO-N, LIGHT IS EMITTED EVERY 5 SECONDS; DURING T-OFF, THE DEVICE REMAINS SWITCHED ON BUT WITHOUT EMITTING LIGHT TO ALLOW THE APPLICATOR TO COOL DOWN FASTER. 42