Safety Guidelines
•
To avoid the risk of electric shock, this equipment must only be connected to a
supply mains with protective earth.
•
In the event of an electrical fire, only use extinguishers that are labelled for that
purpose. Using water or other liquids on an electrical fire can lead to fatal or serious
personal injury.
•
In the event of an electrical fire, to reduce the risk of electrical shock, try to isolate
the equipment from the electric source before attempting to extinguish the fire.
•
Equipment not suitable for use in the presence of a flammable anaesthetic mixture
with air or with nitrous or oxygen enriched atmospheres.
CAUTION
Failure to follow instructions below may result in minor or moderate bodily injury or
damage to the device.
•
If the optional deadman handswitch is installed on the scanner, press and hold
handswitch before pressing the Scan button, and continue to hold handswitch down
for the duration of the exposure (X-ray light on). Early release of the handswitch
stops the exposure and the Fault light turns on. The patient will have to be rescanned.
•
If the scanner or the scanner controller is powered off, at scanner power up or
scanner controller boot up, when the first scan is initiated (scout or full scan), the
i-CAT FLX performs a reset procedure that may delay the start of first scan.
Safety Devices
Warning System
The scanner is equipped with provisions for warning lights and/or audible alarms when X-ray
power is energized.
Interlock System
The scanner is equipped with provisions for an interlock circuit which, when opened, will turn
off X-ray power. Use of the interlock circuit is optional based on site requirements.
The interlock circuit can be used for a door switch, a deadman handswitch, or both. If a door
switch is installed on the room door where the scanner is located, the exposure will stop if the
032-0329-EN Rev E
A-3