4. If using a filter bottom plate, place a large (5.08mm) alignment disk into the selected well; for
other hard bottom plates, no disk or metal ball is required
Note: Verify that the plate is not warped. Warped plates may lead to incorrect probe height
adjustment.
5. Click Eject
at the bottom of the screen.
6. Place a new, empty plate onto the plate holder with well A1 in the top left hand corner.
7. Insert a new, empty plastic strip well into the selected lane and place the reservoir into the
reservoir holder.
Note: Verify that all selected wells and reservoirs are EMPTY and DRY.
8. Click Retract
.
9. Click Auto Adjust Height
LABScan3D and xPONENT 4.2 Quick Reference Guide
. The height will automatically save to the Plate Name.