12 SECTION 12 Procedure to Disconnect the Analyzer When you disconnect the analyzer (e.g., for maintenance), follow the steps below: 1. Make the last measurement with demineralized water, the reading will be equal to zero. 2. Remove the standard reagent bottle and replace with deionized water. Activate the feature fill hoses (MENU> REAGENTS> HOSE) repeat a second time. 3. Install funnel with deionized water, loosen the screws from the peristaltic pump and observe the drain (which should have stopped) without draining the water from the funnel. 4. Leave the unit on STANDBY and watch the drain, which should be discarding the water through the funnel. 2395 Phosphate Analyzer Thermo Scientific Orion | 45