Download IC System Operational and Performance Qualification User Guide
Transcript
IC System OQ/PQ User’s Guide 15. If a gradient pump is to be qualified, select the appropriate Grad Accuracy sequence template and copy it to the Validation directory. If the system contains an isocratic pump go on to Step 16. Note: Microbore (MB) and Standard Bore (SB) pump configurations use different sequences for the Gradient Accuracy Test. GP40 and GP50 microbore pumps should be qualified using the GP40_GP50_MB_Grad_Accuracy sequences. GS50 and standard bore GP50 pumps should be qualified using the GP50_GS50_SB_Grad_Accuracy sequences. Standard bore GP40 pumps should be qualified using the GP40_SB_Grad_Accuracy sequences. 16. If the system to be qualified contains an ED40 or ED50 Detector used as an Amperometry Detector, select the Int_Amp_Noise_Drift and/or the DC_Amp_Noise_Drift sequence template. Copy the template into the Validation directory. These tests should only be performed if the ED40 or ED50 Detector is used in the Integrated Amperometry or DC Amperometry mode. 17. If the system to be qualified contains and EG40 Eluent Generator, select the EG40 Current Qual sequence template and copy it into the Validation directory. 18. Lastly, copy the report file into the Validation directory. Select the OQ_UV.rdf or the PQ_UV.rdf for qualification with an AD20, AD25 or PDA-100 detector. Select the OQ_ECD.rdf or the PQ_ECD.rdf file for qualification with a Conductivity Detector. 19. Open the Validation directory. Right click on the first sequence template and select Properties. Select the timebase that will be qualified in the Timebase entry field. In the Preferred Report and Channel section, select the report (rdf) file newly copied into the directory of the timebase. Confirm that the correct detector channel is selected. 20. Repeat step 19 for all other OQ/PQ sequence templates in the Validation directory. DOC. 031726-02 11/00 Page 16 of 84