Read out the result.
Note:
- For R <10Ω there is a continuous beep and LED lights green.
- Remarks and messages are the same as in par. 3.2.1.
3.2.3 Calibration of test leads
In order to eliminate the impact of the resistance of test leads on measurement result (RCONT and
RX), the compensation (auto-zeroing) of resistance may be performed.
Set the rotary switch of function selection at
RZERO position.
Short the test leads
– READY message
should be displayed.
Press START.
AUTO-ZERO message appears
that confirms completion of test
leads calibration.
The result is a compensated
value and the amendment is
available for RCONT and RX. The
compensation is active even after
the meter is switched off and on
again.
OPERATING MANUAL MIC-2510 version 1.09