Download manual 1 - Digitimer
Transcript
INSTRUTECH CORPORATION VR-10B USER’S MANUAL 8. Trigger segment: This is a four digit hexadecimal number ranging from one to six which sets the protocol segment that the sync trigger pulse will be timed from. When the respective segment is started the trigger output delay takes place, and then the sync trigger pulse is sent. This number must be less than or equal to the number of segments in use. 9. Number of segments: This is a four digit hexadecimal number ranging from one to six that indicates the number of segments that are to be used in the stim protocol. 10. Channel two trigger: This is a four digit hexadecimal number that is broken down into binary bits. The high order eight bits are always zero, as is bit zero. Bits one through six correspond to the segments in the stim protocol. Bit one corresponds to segment one and bit six corresponds to segment six. If this bit is set to one, the trigger two output will be active when the given segment is active. Bit seven indicates an output inversion. If this bit is zero the trigger output will go high when the trigger output is active. If this bit is one, then the trigger output will go low when active. Example: sending 0086h to the VR-10B will cause the trigger two output to go low when segments one and two are active. 11. Segment Descriptions: Parameters 11a through 11f are to be entered for each segment (up to six, as specified in the number of segments parameter, number nine above). They are entered in series with parameters 11a through 11f for segment one followed by parameters 11a through 11f for segment two, etc. 11a. Initial Amplitude: This is a four digit signed hexadecimal number from 8000h to 7FFFh. It specifies the amplitude of this segment at the start of a stim series. The increment value for this parameter is set in the same fashion as number six, the holding potential, however the reference point in this case is not zero, it is the holding potential. Example: if the holding potential is one volt, and a one volt above holding potential pulse is desired, this value is set to one volt. If it is desired for this pulse to go to zero volts then this value must be set to minus one volt. 11b. Amplitude Change: This is a four digit signed hexadecimal number indicating the value to increase or decrease the amplitude of this segment during its execution in each series. This value is set as in 11a. Example: if this parameter is set to one millivolt, the initial amplitude is set to ten millivolts, and the number of cycles is set to ten, then the first cycle would have an amplitude of ten millivolts, the second would be eleven millivolts, the third would be twelve millivolts, this would continue until the tenth cycle is nineteen millivolts. 42