INSTRUTECH CORPORATION
VR-10B USER’S MANUAL
CAL CHK 1CH- IN1 -> OUT1. CUE next. is the next message. This is
similar to the previous step, except that this is to check the settings for the channel
one single channel mode filter. A signal applied to the channel one analog input
is reproduced on the channel one analog output. Gain and zero settings should be
verified. Depress the CUE/FRAME button after this is completed.
At this point you must select the video format of the video recorder being used.
The message “PAL - press CUE, NTSC press FRAME RESET” will be
displayed. At this time select the NTSC (American) video format or PAL
(European) video format by pressing the appropriate button.
CAL OVER - FRAME RESET to save, CUE to abort. is displayed next.
Depress the FRAME RESET button to save the calibration settings just entered.
Depress the CUE/FRAME button to abort these settings and start the procedure
over again. If the FRAME RESET button is not depressed, the calibration
settings just entered will not be saved, and will be lost. If the FRAME RESET
button is pressed, the display will show some random characters briefly, then
display Cal. values saved. CUE to restart. If the CUE/FRAME button is
pressed to abort the calibration then the display will scroll Cal. values not
saved!!! CUE to restart. At this time turn the VR-10B off and remove the
jumper from the calibration pins. The VR-10B should reset when the power is
turned on and go into the normal record mode.
This completes the recalibration of the VR-10B. The VR-10B should now have a
maximum input range of ±10 volts. If this range is exceeded, the over range light
for the particular channel will light, and if the signal is observed on the output
with an oscilloscope, clipping will be apparent.
The video format of the VR-10B will be the format selected in the video format
calibration step. This format must match the format that the VCR is designed for.
The PAL and NTSC standards are not compatible in any way.
If there are any questions or difficulties with this procedure, please contact
Instrutech Corporation for assistance.
9.3 LC Display Contrast
The contrast of the LCD on the VR-10B can be adjusted. Remove the top cover as
described in the recalibration procedure above. On the front left hand side of
control labeled LCD ADJ. This control is located immediately behind the display
assembly. Use care not to adjust the wrong control. Adjust this control while
viewing the display from the normal viewing angle for best contrast.
51