Download TMO2D Oxygen Analyzer - GE Measurement & Control
Transcript
9/30/93 No Compensation (cont.) Damped Response: Set System Response: normal [DAMPED] At this prompt, use the NO or arrow keys to select “DAMPED,” then press YES to confirm the entry. Calibrate - Oxygen Enter Grid Data? Press YES to enter grid data. O2 Grid Entry # of Points [ 3 ]: Use the numeric keys to enter the number of points in the gas calibration curve, then press YES twice to confirm the entry and proceed to the next prompt. Point #1 %O2 [ XX.XX ]: Use the numeric keys to enter the percent gas concentrations for each of the points in the curve as prompted, then press YES to confirm each entry. Note: Percent gas concentrations MUST be in ascending order. XX.XX %O2 O2 mA [ XX.XX ]: Use the numeric keys to enter the corresponding mA values for the given percent gas concentration. The unit will then display the “Working. . .” message before returning to the Calibrate - Oxygen option. Working . . . 4-38 Calibrate - Oxygen Enter Grid Data? At this prompt, press NO to proceed to the next option. Calibrate - Oxygen Done? Then press YES to exit to the Calibration Menu. Programming the TMO2D