The Start Point Parameter About the Start The Start Point is the raw data point where you want base calling to Point start in the sample file. The Start Point is normally the same as the beginning of the first base peak (the Peak 1 Location value). Changing the Start If any of the raw data immediately after the Peak 1 Location is clearly Point unusable, or if you want to analyze only a portion of the raw data, you can start calling bases later in the raw data. In such a case, the Start Point value is greater than that of the Peak 1 Location value. The Start Point value can never be less than the Peak 1 Location value. IMPORTANT If you want to start analysis further along than the actual location of the first base peak, change the Start Point value, not the Peak 1 Location value. Changing the Peak 1 Location value affects the way the DyeSet/Primer file is applied to correct for mobility shifts. Making the To have the Sequencing Analysis software recalculate the Peak 1 Software Location, Start Point, or Stop Point after you have changed a setting, Recalculate enter a zero in the Peak 1 Location field and reanalyze the data. 6-16 The Processing Parameters