Download User Manual
Transcript
12-Channel Electrocardiograph User Manual Operation Instructions for Exercise ECG (Optional) communication protocol Select Baudrate and set this to 4800 Baud Press the red button to leave the service menu Turn off the Valiant 3) For Lode ergometer Turn on the ergometer By default, the Corival ergometer is shipped with the LODE RS232 communication protocol. It should be set to the ERGOLINE P4 protocol. You can change the RS232 protocol by following the procedures below: • Turn on the ergometer and press Enter • You will be in MAINMENU • Select, by using the Up/Down keys, the menu SYSTEM PARAMETER and press Enter • Select, by using the Up/Down keys, the menu SETTINGS and press Enter • Select, by using the Up/Down keys, the menu RS232 PROTOCOL and press Enter • Select, by using the Up/Down keys, ERGOLINE P4 and press Enter • Save the selected RS232 protocol or restore the default RS232 protocol, which is also the LODE PROTOCOL. • Turn off the ergometer For controlling the ergometer by SE-12 Express, the ergometer should be set in the ANALOG mode. The ANALOG mode should be selected as follows. After turning on the ergometer, press Enter, select ANALOG and press Enter again. NOTE: You can also let the ergometer start up in the ANALOG mode every time you turn on the ergometer, by using the default start menu. To set the ergometer for starting up in the ANALOG mode, proceed as follows: • Turn on the ergometer and press Enter • You will be in MAINMENU • Select, by using the Up/Down keys, the menu SYSTEM PARAMETER and press Enter • Select, by using the Up/Down keys, the menu SETTINGS and press Enter • Press Enter to confirm the menu DEFAULTSTARTMENU and press Enter If you turn off the ergometer, it will start up in the ANALOG mode every time the ergometer is turned on again. The Lode ergometer is now ready for external control For details on setting other models for external control, please refer to the connection instruction of the models. - 162 -