Performance Test 4 Table 4-3 DC current verification test CAUTION Function Input Range Error from nominal one year DC current 0.00 μA 500 μA ±0.05 μA 0.0000 mA 5 mA ±0.4 μA 0.000 mA 50 mA ±4 μA 0.00 mA 500 mA ±40 μA 450.00 μA 500 μA ±0.27 μA 4.5000 mA 5 mA ±2.6 μA 45.000 mA 50 mA ±26 μA 450.00 mA 500 mA ±0.26 mA 0.0000 A 5A ±0.5 mA 0.000 A 10 A ±5 mA 4.5000 A 5A ±11.7 mA 9.000 A 10 A ±27 mA Connect calibrator to multimeter’s A and Lo terminals before applying 10 A. U3401A User’s and Service Guide 61