Download Druck DPI 145 Multi-Function Pressure Indicator User Manual K147
Transcript
3: Installation 3-13 To connect an external transducer, plug the connector into the appropriate socket. Resonant sensors should be plugged into the socket labelled EXT RPT and Piezo Resistive sensors into the socket labelled EXT PDCR. NOTE: The external transducer input sockets, although of the same pattern, are electrically different i.e. a Piezo Resistive device will not operate if plugged into the Resonant Sensor socket and similarly, vice versa. Secure the connector into the socket by tightening the connector securing screws. For information on the use of external transducers with the equipment, refer to Section 4.6.1. Section 5, Calibration, gives details of the calibration procedures for external transducers. When a conventional PRT (not Digitally Characterised) is connected to the instrument for the first time, the pressure reading will be shown on the display in units of voltage not pressure. This is because the zero and full scale offsets of the transducer must be corrected by compensation before voltage can be interpolated as pressure. Refer to Section 5 for compensation details. 3.7 Switch Test and Analogue (I/V) Output Connections to the Switch Test and the I/V Output sockets are made via coaxial BNC connectors. To connect to either of these sockets, plug a lead fitted with a BNC connector into the socket. The Pin-out of both sockets is shown in Fig. 3.9. I/V Output Shell Centre Switch (-ve) (+ve) Shell Centre Switch -ve Switch +ve +5V 200 Voltage or Current Source + GND EXT Load Current or Voltage Output +ve + External Switch GND -ve Switch Test Input Figure 3.9 - Switch Test and I/V Output Socket Connections K147 Issue No. 4 3