Download instruction-manual-r..
Transcript
6.0 6.1 APPENDICES Specifications General Water in Oil range Minimum sample flow rate Resolution Temperature stability Temperature sensor Temperature compensation range Shipping weight 0.00 to 50.00% 4 USGPM for oil (no wax) 9 USGPM for condensate 0.001% water 0.0015% water/°C 1000 Ohm platinum RTD 32°F-300°F (0°C-150°C) Approx. 40 Lbs (Varies by model) Mechanical Construction material Maximum working pressure Maximum fluid temperature Maximum ambient temperature Operating temp for electronics Storage temp for electronics Sample line connection size Specific volume Sensor cross section flow area Type 316 SS 2160 psig @ 100°F (15 MPa @ 38°C) 257°F (125°C) 185°F (85°C) -40°F to +185°F (-40°C to +85°C) -40°F to +185°F (-40°C to +85°C) 1" NPT or 1”and 2” flanged 0.070 U.S. gallon (235 ml) Approx. 0.75 inch2 (4.84cm2) Electrical Power Required Enclosure 18 - 28 VDC @ 150 mA. (Grounded or ungrounded) Hazardous area Class I, Div. 1, Gr. C&D EEx d IIB T4, zone 1, IP65 Output Analog current Maximum load Isolation voltage Relay contacts Adjustable delay Adjustable hysteresis 4 - 20 mA. DC, isolated, self-powered 600 Ohms 500 Volts peak SPST-NO 2 Amp @ 24V DC, Non-inductive 0-43 million seconds 0-49.90% water Terminal/Computer Interface. RS-232C Speed Word size Parity Stop bits Full duplex 9600 Baud 8 Bits None One RS-485 Half duplex (application specific) - 35 - ©2004, Delta-C Technologies Inc.