POD-201 User Guide Custom Calibration The calibration setup window allows the user to select the polarimeter calibration matrix to be used for SOP measurement. Factory Matrix: The POD-201 or PSY-201 comes with a set of factory-generated calibration matrices for wavelengths in its operating wavelength range. If “Factory Matrix” is selected, then the wavelength setting determines the built-in calibration matrix that is used for measurement. Factory matrices are spaced at 5 nm intervals for 1550nm instruments and at 2nm intervals for 1310nm instruments. Custom Matrix: If this option is selected, the user can load a user-generated calibration matrix to use for measurement. Matrix Calibration: Use this function to generate a new calibration matrix for the current operating conditions (wavelength, temperature). Such a calibration matrix can yield more precise results under particular operating conditions, or can be used to generate calibration matrices for wavelengths outside of the instrument’s nominal operating range, although the measurement performance may not be as good as for wavelengths in the range. To generate a new matrix, type the light source wavelength into the “Wavelength” box, and click “Calibrate GP-UM-POD-201-11 Page 71 of 83