Download 1300 Series A2 BSC - Operating Manual

Transcript
Section 7
Specifications
Volume, weights, and loads
3ft Models (10” window opening work
position) - 1320, 1321, 1322, 1323, 1330,
1331, 1332, 1333
3ft Models (8” window opening work
position) - 1339, 1340, 1341, 1342, 1390,
1391, 1392, 1393
5ft Models (10” window opening work
position) - 1368, 1369, 1370, 1371, 1380,
1381, 1382, 1383
5ft Models (8” window opening work
position) - 1350, 1351, 1352, 1353, 1366,
1367, 1372, 1373
l / gal
15 / 4.0
25 / 6.6
Cabinet
kg / lb
170 / 375
230 / 507
Stand
kg / lb
24.4 / 53.8
28.1 / 62.0
Description
Volume
Drain pan
Weights
Loads
Max. load on single work surface module kg / lb
Maximum load on overall working area
kg / lb
25 / 55
50 / 110
75 / 165
Ambient conditions (indoor use only)
Temperature
Max. ambient temp during operation
°C / °F
40 / 104
Min. ambient temp during operation
°C / °F
10 / 50
Max. humidity / operation
% r.H.
90, non-condensing
Max. humidity / storage
% r.H.
95
Humidity
Heat dissipation to environment
Room temperature 20°C
kJ/s / BTU/s
3ft models 10": 0.17 / 0.16
3ft. models 8": 0.15 / 0.14
5ft. models 10": 0.31 / 0.29
5ft. models 8": 0.28 / 0.27
<2
<2
3ft. models 10”: 63, ±2
3ft. models 8": 62, ±2
5ft models 10”: 65, ±2
5ft. models 8": 64, ±2
Room temperature rise
Above room temp with window closed
°K
Ergonomics
Noise level*
dB(A)
*The noise level was determined in accordance with NSF/ANSI 49. The measurement uncertainty is within a range of ± 2 dB.
7-4
1300 Series A2
Thermo Scientific