OPERATION
Intelect® Legend XT Therapy System
PATIENT DATA CARD SETTING UP OF NEW CARD (continued)
Setting up Electrodes
Accessing Electrode Placement
Press the Pad button to select the Red (+) or
The following information uses an IFC
Black (-) electrode.
Treatment as an example. Electrode
Placement procedures for all modalities are
Press the Chan button to select Channel 1
performed in the same basic fashion.
or Channel 2.
Press the Electrode Placement button.
ELECTRODE PLACEMENT BUTTON
Press the Side button to select Front, Back,
Left, or Right of the body graphic.
Press the Size button until desired electrode
size is displayed. If electrode desired is not
listed, select Other.
PAD
BUTTON
Electrode Placement
Press the Up, Down, Left and Right Arrow
buttons to position the selected electrode
as close to the actual treatment location as
possible.
Press the Pad button to select the other
electrode. Repeat above procedure for
electrode positioning.
If applicable, press the Chan button, to
select another channel and repeat above
procedures.
After positioning the electrodes, press the
Accept and Return Arrow button.
ELECTRODES
POSITIONED
CHAN
BUTTON
SIDE
BUTTON
SIZE
BUTTON
NOTE: When Ultrasound is the modality,
only the Side button is available.