Download IJ Instruments Ltd. IJ-6 SOFTWARE
Transcript
11.2 11.0 Voltage (V) 10.8 10.6 10.4 10.2 10.0 9.8 0 10 20 30 40 50 Currrent (mA) Fig. 17: SSR output voltage and current Fig. 18 shows an example of connecting the temperature controller to a “brick-type” solid-state relay. A reed relay or optoisolator can also be connected instead, providing the current draw is kept below the maximum of 40mA. Solid-state relay Observe correct polarity! To/from switched load + Fig. 18: Connections for external solid-state relay 8.2.3 THERMOCOUPLE GROUND REFERENCE A suitable thermocouple ground reference must be chosen, depending on whether the thermocouple is floating or grounded. The two rightmost terminals of the terminal block, marked “0V” and “-”, are used for this purpose. “0V” is connected to the negative supply rail of the temperature controller and hence to the negative supply of the USB port. “-” is connected directly to the negative terminal of the thermocouple connector. Depending on how these terminals are connected, either a floating or grounded thermocouple can be used, as described below. If the thermocouple is • • bare (e.g. welded-bead) and not attached to anything (e.g. placed in mid-air to measure air temperature) bare and attached to an object, but the object itself is floating (not grounded) then the thermocouple is “floating” relative to ground and the connection shown in Fig. 19 should be used. A shorting link is placed between “0V” and “-”, thus connecting the negative terminal of the thermocouple to 0V. The maximum allowed floating voltage is +/- 3V relative to ground. 20