Download Operator`s Manual
Transcript
-------------------------------------------------------------------- HCP-7000 Opreator’s Manual 9 The Model HCP-7000 system is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The user of the HCP-7000 system can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the HCP-7000 system as recommended below, according to the maximum output power of the communications equipment. Separation distance (m) according to frequency of transmitter Rated maximum output power (W) of transmitter 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 1.0 GHz 0.01 0.12 0.12 0.23 0.1 0.37 0.37 0.74 1 1.17 1.17 2.33 10 3.70 3.70 7.37 100 11.70 11.70 23.30 For transmitters rated at a maximum output power not listed above, the recommended separation distance (d) in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people. Immunity and Compliance Level Immunity test IEC 60601 Test Level Conducted RF IEC 3 Vrms 150 kHz to 80 61000-4-6 MHz Radiated RF IEC 61000- 3 V/m 80 MHz to 1.0 4-3 GHz Actual Immunity Level Compliance Level 3 Vrms 3 Vrms 3 V/m 3 V/m