Download Figure 1.

Transcript
Command
Control Form
[:SOURce]:WAVElength:SLEWrate:FORWard<wsp><value>
Query Form
[:SOURce]:WAVElength:SLEWrate:FORWard?
Description
Write and read the forward slew rate for a wavelength scan.
Explanation:
Control Form
This command does not affect and is not affected by whether the laser is in
local mode or remote mode.
1 <= VALUE <= 100 (The units are in percentage of the maximum rate.)
If the value is out of range:
• Returns Out of Range.
• No other action is taken.
If the value is in range:
• Returns OK.
• Forward wavelength slew rate is set to the desired value.
• The 1 to 100% value is translated into a voltage on the picomotor control
board. The actual speed is not linear with this voltage. As you increase
the slew rate from 1 to 100% the actual wavelength rate of change does
not always increase.
• The slew rate only affects scanning. The slew rate during a wavelength
change from a SOURCE:WAVELENGTH command is always 100%.
Explanation:
Query Form
?
Returns the current forward slew rate.