Download Manual - Astronics Test Systems
Transcript
1256E User Manual PROBLEM Publication Number 980873 Rev. A POSSIBLE CAUSES • Baud rate too high for cable and terminal being used. • Cable is excessively long, or not of high quality. Text from 1256 to remote terminal overwrites itself, staying on one line. • Termination characters of 1256 not compatible with terminal setup. Text from 1256 to remote • Termination characters of terminal is double-spaced 1256 not compatible with (lines of text are separated by terminal setup. blank lines). Remote terminal does not display characters that are typed. • Echo mode of remote terminal is not set to halfduplex (local echo). • HyperTerminal (included with Microsoft Windows) can fail to echo characters properly. Troubleshooting C-4 WHAT TO DO • Change to a lower baud rate. Make sure that the new 1256 baud rate matches the new terminal baud rate (to set the 1256 baud rate, see Chapter 4, GPIB and RS-232 Interfaces, under the heading Setting the Baud Rate). • Use a high-quality, shielded cable. • Set 1256 to use carriage return and line feed (CR-LF) as termination characters (see Chapter 4, GPIB and RS-232 Interfaces, under the heading Setting the Message Termination Characters). • Set 1256 to use only carriage return (CR) as termination character (see Chapter 4, GPIB and RS-232 Interfaces, under the heading Setting the Message Termination Characters). • Set terminal for half-duplex (local echo) mode. Consult the terminal user’s manual for instructions. • Download newer version, known as “HyperTerminal Private Edition” from: http://www.hilgraeve.com Astronics Test Systems