Download Manual - Astronics Test Systems
Transcript
Publication Number 980873 Rev. A • 1256E User Manual Most private networks found inside commercial/industrial environments are typically contiguous which means that one instrument can communicate directly to other computers and equipment without the assistance of an intermediate gateway. In this situation, the gateway address need not be specified in the 1256. If the 1256 exists in a network that has individual subnets isolated by gateways or bridges, the gateway parameters must also be specified or the 1256 will only be able to communicate with computers and equipment local to its own subnet. To select manual mode, proceed as follows: 1. Press the top key, which will select line 1 (top line). Rotate the knob until the top line reads: >-----SETTINGS-----< 2. Select line 2 (next to top line) by pressing the second key. Rotate the knob until the display shows line 2 as follows (lines 3 and 4 may differ): -----SETTINGS----> Remote: TCP/IP < Option: IP_Oct_1 Status: 0 3. Press the third key, selecting line 3. Rotate the knob until line 3 reads as follows (line 4 may differ): -----SETTINGS----Remote: TCP/IP > Option: Addr_Mode< Status: Auto_DHCP 4. If line 4 already indicates the Manual Set mode, no further action is necessary. Otherwise, select line 4. Rotate the knob until the display reads: -----SETTINGS----Type: TCP/IP Option: Addr_Mode <Update?: Man_Set Then press key 4 again to select the Manual Set mode. This temporarily disables the TCP/IP interface, allowing you to change the TCP/IP settings. Astronics Test Systems Computer Interfaces 4-7