Download Printable Version

Transcript
Selecting default settings
AUTO
LEVEL&TIME
FS-K5
The AUTO LEVEL & TIME softkey automatically sets the trigger offset and the
reference level.
If necessary, the IF or RF trigger level and the setting of the mechanical and (if
available) electronic attenuator are optimized in addition.
If the measurement is not completed successfully (termination by user or as a
result of an error), no setting is changed.
The following parameters must be correctly set before this auto function is
started:
- center frequency
- GSM trigger mode
- modulation mode
- multislot settings
- midamble
If automatic setting is not successful, make the settings manually with the
REF LEVEL and TRIGGER OFFSET softkeys.
IEC/IEEE bus command
POINT/SYMB
MULTISLOT
4
8
READ:AUTO:LEVTime?
The POINT/SYMB softkey switches the resolution for
measurements between 4 and 8 points per analyzed symbol.
PFE/PVT/MAC
This settings does not influence the CPW/MOD/TRA/SPU measurements.
For POINT/SYMB = 8, twice the amount of data is processed. As a result, the
resolution for the PFE/MAC/PVT measurements increases. The measurement
speed, however, decreases because of longer computing times. The data can
be retrieved by using the TRACE ASCII EXPORT function of the basic unit or by
using the remote control command TRAC:DATA? TRACE1.
IEC/IEEE-bus command:
MODULATION
MULTISLOT
GSM EDGE
CONFigure:PRATe 4 | 8
The MODULATION GSM / EDGE softkey selects the modulation type. The limit
lines are adapted to the selected modulation type upon switchover.
IEC/IEEE bus command
1141.1515.44
CONFigure:MTYPe GMSK | EDGE
2.12
E-5