Download Operator
Transcript
ORTEC ® Detective-EX/DX ®-100T Portable HPGe Spectrometer — User’s Manual NOTE 931040P / 1014 This section discusses only the mandatory background measurement required to complete the manual calibration. For the full discussion of Detective-EX background measurements, see Section 5.2.2. As noted elsewhere, until you complete the background measurement, the Identify and Monitor Modes will be disabled. The minimum acceptable data collection time for the background meas is set on the Bkg Settings screen (see Fig. 61 on page 60). Note that, for optimum counting statistics, we recommend a minimum 30-minute background acquisition. However, the background settings screen also allows minimum intervals of 1, 5, or 15 minutes. ! All sources should be removed from the area before you tap the Run a Background button at the end of the manual calibration routine. If you leave the source in place, a “high count rate” message will remind you to remove the source and restart the calibration. ! Tapping the Run a Background button opens the screen shown in Fig. 22 and automatically starts the background acquisition. ! Acquisition continues until you tap either Save or Cancel, which returns you to the Advanced menu. If the measurement is canceled before that minimum time has elapsed, the software does not accept (i.e., save and use) the new background spectrum. ! If necessary, you can obtain a valid background later. See the instructions in Section 5.2.2. Figure 22. Collect Background for At Least the Specified Minimum Time. 26