4. SETTING PARAMETERS
Wiring Diagram
Input (RTD)
A
B
B
8
4
7
7
4
8
Power supply
L+
N−
GND
14
15
The parameters are set as you specified in your order. Refer
to the following to change the default settings.
Set the parameters using a PC (VJ77 Parameter Setting Tool)
or the Handy Terminal. Refer to "5. List of Parameters" in this
manual and the User’s Manual for VJ77 PC-based Parameters
Setting Tool (IM 77J01J77-01E) or the User’s Manual for JHT200
Handy Terminal (IM JF81-02E). Parameters are shown in
brackets "[ ]".
Output 2 (opt.)
9 +
10 −
Output 1
9
10 11 12
14 15 16
11 +
12 −
Connection of Setting Tools
16
JHT200
Handy Terminal
JUXTA communication cable
3 pin connector (F9182ED)
[Provided with VJ77 and JHT200]
NOTE
●Do not connect anything to the terminals that are not
used in the wiring diagram. Otherwise it may cause
the malfunction or damage.
●The power line and input/output signal lines should
be installed away from noise-generating sources.
Otherwise accuracy cannot be guaranteed.
●The wiring resistance of input terminals
and
should be the same.
●The grounding resistance must be 100Ω (JIS Class
D grounding). The length and thickness of the
grounding cable should be as short and thick as
possible. Directly connect the lead from the ground
terminal
of the product to the ground. Do not
carry out daisy chained inter-ground terminal wiring.
●The ground terminal
becomes effective for AC
power only.
●Adhere strictly to the specifications to avoid
overheating or damage. Before turning on the power,
ensure the following:
(a) Power supply voltage and input signal value
applied to the product should meet the required
specifications.
(b) The external wiring to the terminals and wiring to
ground are as specifications.
●Do not operate the product in the presence of
flammable or explosive gases or vapors.
●The product is sensitive to static electricity; exercise
care when in operation. Before you operate the
product, touch a nearby metal part to discharge
static electricity.
VJ77 Dedicated adapter (E9789HA)
[Provided with VJ77]
Dedicated cable (E9786WK)
[Provided with VJ77]
PC with VJ77 installed
*Be sure to use the VJ77 of version R1.05 or later.
Setting RTD Type
Set the RTD type to be connected to the input in
[B05: TYPE].
Display
Type of input
PT
Measuring
range
-200 to 660°C
Pt100
(IPTS-68:JIS'89)
JPT
JPt100 (JIS'89)
-200 to 10°C
PT100-90 Pt100 (ITS-90:JIS'97) -200 to 850°C
PT50
Pt50 (JIS'81)
-200 to 649°C
Measuring
span
10°C
or more
JPt100 (JIS'89): R0=100Ω, R100/R0=1.3916
Pt100 (IPTS-68): R0=100Ω, R100/R0=1.3850
Pt100 (ITS-90): R0=100Ω, R100/R0=1.3851
Setting Temperature Unit
Set the temperature unit of the input range in [B09: UNIT]
NOTE
Do not set the "-----" as the temperature unit, even if it
is displayed during the setting.
Otherwise it may cause the malfunction or damage.
Power Supply and Isolation
Power supply voltage:
100 to 240V AC (-15%, +10%) 50/60Hz
24V DC ±10% (ripple: less than 5% p-p)
Current consumption
DC: WR5A 65mA, WR5V 50mA
Power consumption:
100V AC: WR5A 4VA, WR5V 3VA
200V AC: WR5A 5.5VA, WR5V 4.5VA
Insulation resistance: 100 MΩ at 500 V DC between input,
output, power supply, and grounding terminals
mutually.
Withstand voltage:
DC: 1500V AC/minute between input-output, inputpower supply
500V AC/minute between output-power supply
AC: 1500V AC/minute between input, output, power
supply, and grounding terminals mutually.
Setting Input Range
Set the zero percent value of input range in [B10: ZERO], and
the span of input range in [B11: SPAN].
Setting Burnout Action
Set the burnout action in [B12: BURN OUT]
Set "OFF", "UP", or "DOWN."
NOTE
Input adjustment should be applied when the setting
of input range, burnout action direction or input wiring
has been changed.
IM 77J09R05-01E