Download User`s Manual
Transcript
4. SETTING PARAMETERS Wiring Diagram Input (RTD) A B B 8 4 7 7 4 8 Power supply L+ N− GND 14 15 The parameters are set as you specified in your order. Refer to the following to change the default settings. Set the parameters using a PC (VJ77 Parameter Setting Tool) or the Handy Terminal. Refer to "5. List of Parameters" in this manual and the User’s Manual for VJ77 PC-based Parameters Setting Tool (IM 77J01J77-01E) or the User’s Manual for JHT200 Handy Terminal (IM JF81-02E). Parameters are shown in brackets "[ ]". Output 2 (opt.) 9 + 10 − Output 1 9 10 11 12 14 15 16 11 + 12 − Connection of Setting Tools 16 JHT200 Handy Terminal JUXTA communication cable 3 pin connector (F9182ED) [Provided with VJ77 and JHT200] NOTE ●Do not connect anything to the terminals that are not used in the wiring diagram. Otherwise it may cause the malfunction or damage. ●The power line and input/output signal lines should be installed away from noise-generating sources. Otherwise accuracy cannot be guaranteed. ●The wiring resistance of input terminals and should be the same. ●The grounding resistance must be 100Ω (JIS Class D grounding). The length and thickness of the grounding cable should be as short and thick as possible. Directly connect the lead from the ground terminal of the product to the ground. Do not carry out daisy chained inter-ground terminal wiring. ●The ground terminal becomes effective for AC power only. ●Adhere strictly to the specifications to avoid overheating or damage. Before turning on the power, ensure the following: (a) Power supply voltage and input signal value applied to the product should meet the required specifications. (b) The external wiring to the terminals and wiring to ground are as specifications. ●Do not operate the product in the presence of flammable or explosive gases or vapors. ●The product is sensitive to static electricity; exercise care when in operation. Before you operate the product, touch a nearby metal part to discharge static electricity. VJ77 Dedicated adapter (E9789HA) [Provided with VJ77] Dedicated cable (E9786WK) [Provided with VJ77] PC with VJ77 installed *Be sure to use the VJ77 of version R1.05 or later. Setting RTD Type Set the RTD type to be connected to the input in [B05: TYPE]. Display Type of input PT Measuring range -200 to 660°C Pt100 (IPTS-68:JIS'89) JPT JPt100 (JIS'89) -200 to 10°C PT100-90 Pt100 (ITS-90:JIS'97) -200 to 850°C PT50 Pt50 (JIS'81) -200 to 649°C Measuring span 10°C or more JPt100 (JIS'89): R0=100Ω, R100/R0=1.3916 Pt100 (IPTS-68): R0=100Ω, R100/R0=1.3850 Pt100 (ITS-90): R0=100Ω, R100/R0=1.3851 Setting Temperature Unit Set the temperature unit of the input range in [B09: UNIT] NOTE Do not set the "-----" as the temperature unit, even if it is displayed during the setting. Otherwise it may cause the malfunction or damage. Power Supply and Isolation Power supply voltage: 100 to 240V AC (-15%, +10%) 50/60Hz 24V DC ±10% (ripple: less than 5% p-p) Current consumption DC: WR5A 65mA, WR5V 50mA Power consumption: 100V AC: WR5A 4VA, WR5V 3VA 200V AC: WR5A 5.5VA, WR5V 4.5VA Insulation resistance: 100 MΩ at 500 V DC between input, output, power supply, and grounding terminals mutually. Withstand voltage: DC: 1500V AC/minute between input-output, inputpower supply 500V AC/minute between output-power supply AC: 1500V AC/minute between input, output, power supply, and grounding terminals mutually. Setting Input Range Set the zero percent value of input range in [B10: ZERO], and the span of input range in [B11: SPAN]. Setting Burnout Action Set the burnout action in [B12: BURN OUT] Set "OFF", "UP", or "DOWN." NOTE Input adjustment should be applied when the setting of input range, burnout action direction or input wiring has been changed. IM 77J09R05-01E 1st Edition 2007.01.15-00