5.4
Removing the Flow Cell from the Optical Bench
The flow cell must be removed from the optical bench in order to inspect the
cell for leaks or obstructions, replace the cell windows (see Section 5.5), or
install a new cell (see Section 5.6).
1. Grasp the Dionex PDA front cover by the sides and pull straight off to
remove.
2. Disconnect the tubing connected to the flow cell inlet and outlet (see
Figure 5-1).
3. Remove the flow cell cover from the inside front panel (see Figure 2-3).
4. Standard or semi-micro cell: Remove the tubing bracket from the lamp
cover (see Figure 5-1).
5. Remove the flow cell inlet tubing from the tubing guide in the lamp cover (see
Figure 5-1).
6. Squeeze the handle of the flow cell and pull it out of the optical bench.
Figure 5-1. Removing the Flow Cell from the Optical Bench (Flow cell cover not shown)