Dionex PDA Operator’s Manual UV lamp error. UV lamp error or control circuitry error while lamp is on. UV lamp failed to turn on for the required operation. UV lamp error while attempting to turn on. Visible lamp error. Visible lamp error or control circuitry error while lamp is on. Visible lamp error while attempting to turn on. Visible lamp failed to turn on for the required operation. To troubleshoot: 1. Check that the lamp referred to in the error message is turned on. When a lamp is on, the corresponding LED (DEUTERIUM – UV or TUNGSTEN – VIS) is lighted. If necessary, turn on the lamp from Chromeleon or the TTL input. 2. If you recently installed a new deuterium lamp, check that the UV lamp connector wires are routed through the wire guide in the optical bench (see Figure 5-3) and are not pinched under the lamp cover. 3. Check the number of hours the lamp has been in service: Open the Chromeleon 7 Command window (see page 31) or the Chromeleon 6.8 Commands dialog box (see page 32), and then select UV. If the error message relates to the deuterium lamp, execute the UVLampAge command. 70 Doc. 065378-01 1/12