Download ICS-Series Photodiode Array Detector Manual
Transcript
4 • Troubleshooting 4.10 No Spectra Collected • Spectral range was not programmed In Chromeleon 7, select a spectral range in the instrument method or on the Dionex PDA ePanel. In Chromeleon 6.8, select a spectral range in the program or on the Dionex PDA Control panel. • Detector and instrument method/program (or ePanel/Control panel) connected to different timebases In Chromeleon 7, add the detector to the same instrument as the instrument method or the Dionex PDA ePanel. In Chromeleon 6.8, add the detector to the same timebase as the program or the Dionex PDA Control panel. 4.11 Low Spectral Resolution • Bunch width too wide Select a bunch width setting of 1 nm, and then check the resolution again. • Bandwidth too wide Select a bandwidth that includes only the absorbing wavelength. • Incorrect reference wavelength Select a reference wavelength where minimum absorbance occurs. Make sure the reference wavelength is included in the wavelength range. • Wavelength range too narrow Make sure the spectrum wavelength range includes both the reference wavelength and reference bandwidth. Make sure the peak absorbance wavelength is not included in the reference bandwidth. • Reference bandwidth too wide Select a narrower reference bandwidth. The reference bandwidth should not overlap the bandwidth. Doc. 065378-01 1/12 83