Download Technical Manual
Transcript
SPEC CPI Preliminary Technical Manual Figure 6.7.10. 6.8 CPI Imaging System Camera Alignment The CPI imaging system camera may need to be realigned if the imaging system goes out of alignment over repeated airborne operation or it is removed from the instrument for servicing. This procedure provides a fundamental overview for aligning and focusing the imaging system camera and lens assembly. CAUTION: The overall quality of the images recorded by the CPI depends on the camera being properly aligned and focused. The alignment procedure should only be performed by personnel who have read and become familiar with this procedure. 1. An alignment pin is included with the CPI for use in centering and aligning the imaging system (Figure 6.8.1). 2. Figure 6.8.2 shows the proper way to remove the cover from the alignment pin. The pin has been machined on a lathe to come to a sharp point at an exact distance into the optical block. Be SPEC Inc. reserves the right to make improvements and changes to the CPI and related software at any time and without notice 151