Download MC142 User Manual
Transcript
POWERTEK MC-142 Multifunction Calibrator 2, 20 V DC voltage ranges The functional scheme is shown in the picture below: There is a DC reference voltage source integrated within the 22bit measurement converter. Its output is fed to the output stage for 2 and 20 V range. Output voltage present at Hi and Lo terminals is sensed by sensing wires. Feedback eliminates the influence of amplifier’s output impedance and of the resistance of wires within the calibrator. 200 V DC voltage range The block diagram functional scheme is shown in the picture below. The scheme is similar to 2 and 20 V ranges. A 240 V power amplifier with electronic fuses is connected to the output of 20 V amplifier. 2 to 200 V AC voltage ranges The functional scheme is shown in the picture below: User Manual v41 71