Download User Manual
Transcript
Chapter 4 | Troubleshooting 25 Diffused Signals Table 4.2 Troubleshooting diffused signals Probable Cause Recommended Action Tissue dries up during processing Keep tissue section moist starting from the pretreatment boiling step by: Adding respective reagents immediately after decanting solution from slides. Limiting tissue exposure to air for too long before adding hybridization reagents. Adding all working reagents onto the slides before moving them to the 40 °C humidified incubator. Insufficient washing in 1X PBS Make sure tissues are washed in 1X PBS twice after protease digestion and twice again after subsequent fixing in 4% formaldehyde. Fast Red substrate not freshly prepared Prepare Fast Red substrate immediately before use. Slides are not dried before mounting Ensure that slide sections are completely dry before mounting (about 20 min). Mounting solution contained alcohol Use DAKO Ultramount or Innovex Advantage mounting media to mount your tissue. Avoid any mounting solution containing alcohol. Poor Cell Morphology Table 4.3 Troubleshooting poor cell morphology Probable Cause Recommended Action Incorrect pretreatment conditions See Optimization Experimental Design Layout on page 29. Sample preparation not fixed properly Make sure that freshly-dissected tissues are fixed in 10% NBF or 4% PFA for 16-24 hours. Section thickness not optimal Make sure tissues are sectioned at 5 ± 1 μm thick. Tissue Detachment from Slide Table 4.4 Troubleshooting tissue detachment from slides Probable Cause Recommended Action Insufficient baking of slides Verify that 30 min baking step was done. It may be necessary to increase baking time to 1 hr. Incorrect pretreatment conditions Perform full assay optimization procedure to determine optimal boiling time and protease digestion time. Improper fixation, reagents, or concentrations Make sure the correct concentration of NBF was used to fix the slides in the respective steps. Temperature of pretreatment condition too high Make sure the temperature is within the tolerance range of 95-100 °C. For fatty, soft tissue such as breast, adjust to 95 °C. Protease treatment is too long or at too high concentration Reduce protease concentration and/or incubation time.