Download AmpFLSTR Yfiler PCR Amplification Kit User`s Manual (PN 4358101C)
Transcript
Chapter 4 Performing Electrophoresis Overview Flowchart Extract and quantify data 310 Electrophoresis Setup Set up the instrument PCR amplify data Perform electrophoresis Set up instrument: - Install and clean pump block - Install or replace capillary - Prepare syringes - Clean electrode - Recalibrate autosampler - Fill buffer reservoir Prepare the samples • Prepare formamide size standard cocktail: - 0.5µL GeneScan® - 500 LIZ® Size Standard® - 24.5µL Hi-Di™ Formamide per sample • Vortex Yes ABI PRISM® 3100/3100-Avant ABI PRISM® 310 Analyzer Matrix file created? Create matrix file Not sure Analyze data 4-28 DRAFT May 25, 2005 9:59 am, 310.fm No Display and check matrix file • Into each 0.2-mL or 0.5-mL sample tube add: - 25µL formamide: size standard mixture - 1.5µL allelic ladder or sample • Heat @ 95˚C for 3' then place on ice for 3' Set up Data Collection Software v3.0 If using GeneScan® Software v3.7.1: • Preheat the instrument • Set up the sample sheet and injection list defaults • Create the sample sheet and injection list Perform electrophoresis Load samples Select injection list If using GeneMapper ID Software v3.2: • Preheat the instrument • Create the sample sheet and injection list Start run AmpFlSTR Yfiler PCR Amplification Kit User’s Manual