Download USER MANUAL - Dyne Systems

Transcript
The faults listed in the following table are detected and reported by the Andros analyzer.
Consult the Andros Product Manual for additional information regarding possible causes
and remedies of each fault.
Fault
Description
in-flow fault
There has been a substantial reduction in inlet-side pressure.
out-flow fault
There may be an occlusion at output of the Andros analyzer.
infrared (IR) signal lost
IR signal is lost or weak.
ambient temperature
out of range
The ambient temperate is outside the range 0 to 50 °C.
system fault
An Andros hardware failure has occurred.
IR source temperature
exceeds 75 °C
IR source temperature is too high. Power down the XM for approximately 1/2 hour to let the Andros cool and
then re-apply power. If the problem persists, contact Dyne Systems.
Warnings
A description of all XM warnings is presented in the following table.
Warning
Description
zero calibration
requested
This warning is periodically generated by the Andros analyzer. A zero calibration procedure should be
carried out as soon as possible as described in Chapter 6 (section 1.1). The Andros issues this warning for
many reasons (which are also listed in Chapter 6 (section 1.1)).
new NOx sensor
required
The full-scale output of the NOx sensor has dropped significantly since it was installed. The NOx sensor
should be replaced as soon as possible.
new O2 sensor
required
The output of the O2 sensor has dropped significantly since it was installed. The O2 sensor should be
replaced as soon as possible.
sample vacuum at
warning level
The sample vacuum has exceeded the warning limit (listed in section 3.2.1). Check the heated sample
line and all filters for obstructions. Clean or replace the filters as needed. Make sure the check valve is fully
open.
previous leak check
failed
The previous leak check of the sample line failed. The XM will still operate; however, there is a risk that the
sampled gases are being diluted with ambient air. Perform another leak test (see Chapter 7 (section 1.2)) as
soon as possible and correct any leaks that exist in the sample line and connections.
34
yneSystems • XM 1000A
D
DPD-003-01 • © Copyright DyneSystems, Inc.
Section 4