Download USER MANUAL - Dyne Systems
Transcript
The faults listed in the following table are detected and reported by the Andros analyzer. Consult the Andros Product Manual for additional information regarding possible causes and remedies of each fault. Fault Description in-flow fault There has been a substantial reduction in inlet-side pressure. out-flow fault There may be an occlusion at output of the Andros analyzer. infrared (IR) signal lost IR signal is lost or weak. ambient temperature out of range The ambient temperate is outside the range 0 to 50 °C. system fault An Andros hardware failure has occurred. IR source temperature exceeds 75 °C IR source temperature is too high. Power down the XM for approximately 1/2 hour to let the Andros cool and then re-apply power. If the problem persists, contact Dyne Systems. Warnings A description of all XM warnings is presented in the following table. Warning Description zero calibration requested This warning is periodically generated by the Andros analyzer. A zero calibration procedure should be carried out as soon as possible as described in Chapter 6 (section 1.1). The Andros issues this warning for many reasons (which are also listed in Chapter 6 (section 1.1)). new NOx sensor required The full-scale output of the NOx sensor has dropped significantly since it was installed. The NOx sensor should be replaced as soon as possible. new O2 sensor required The output of the O2 sensor has dropped significantly since it was installed. The O2 sensor should be replaced as soon as possible. sample vacuum at warning level The sample vacuum has exceeded the warning limit (listed in section 3.2.1). Check the heated sample line and all filters for obstructions. Clean or replace the filters as needed. Make sure the check valve is fully open. previous leak check failed The previous leak check of the sample line failed. The XM will still operate; however, there is a risk that the sampled gases are being diluted with ambient air. Perform another leak test (see Chapter 7 (section 1.2)) as soon as possible and correct any leaks that exist in the sample line and connections. 34 yneSystems • XM 1000A D DPD-003-01 • © Copyright DyneSystems, Inc. Section 4