Transcript
Fig.25 Alarm limits setting range: SpO2: Upper alarm range: 71-100 The default SpO2 upper alarm is 100. Low alarm range: 70-99 The default SpO2 low alarm is 95. PR: Upper alarm range: 31-254 The default PR upper alarm is 120. Low alarm range: 30-253 The default PR low alarm is 50. SYS Upper alarm range: 61-255 The default PR low alarm is 160 Low alarm range: 60-220 The default PR low alarm is 90. DIA Upper alarm range: 31-220 The default PR low alarm is 95 Low alarm range: 30-180 The default PR low alarm is 60. MAP Upper alarm range: 31-220 The default PR low alarm is 110 Low alarm range: 30-200 The default PR low alarm is 60. SpO2/PR/SYS/DIA/MAP switch: ON:Turn on the alarm. Off:Turn off the alarm even though the value is out of limits. Note: 1. When the monitor is turned off, the alarm limits of the latest set are restored, so after the monitor is turned on, please set alarm limits again if necessary. 2. The alarm set should be handled by the qualified personnel. CHAPTER 11 Real Print Set(Optional) The monitor can print data by configured printer such as B6 printer. In the menu 2 screen, press the or the button to select the “ Real Print Set” item, and then press the button to enter into the item. Refer to the Fig.26. Press the or button to select the sub-items to set up. And press the button to enter. Set the selected item by pressing the or button, then press the button to confirm. Press the button to returning to the previous menu. - 31 -