Download User`s Manual
Transcript
PC terminal to set trigger source: TRIGger:SOURce BUS By internal trigger: The internal trigger is the default trigger mode for the Remote Interface Operation. In the internal trigger mode, a trigger signal will be issued whenever the multimeter is in the wait-for-trigger state. To set the internal trigger, use the following command from your PC terminal: TRIGger:SOURce IMMediate 4.2.3 Trigger Setting In M3510A/11A, you can specify a variety of trigger settings including the number of samples per trigger, the number of triggers per event, reading hold, and the trigger delay for your measurements. A. Number of samples on each trigger By default, M3510A/11A takes only one reading on each trigger, but you can instruct the multimeter to take specific number (up to 2000) of readings each time it receives a trigger. The defined input setting is stored in a volatile memory which will be cleared after the meter has been turned off and the default value will be restored. You can set the number of samples on each trigger through front panel or the remote interface. Front Panel Operation 1. Press SHIFT+CONFIG button. 2. Set a defined parameter via the procedure TRIG SYS > TRIG CNT > ENTER. 3. △and ▽ buttons to increase or using ◁ and ▷ buttons to move Set the desired number by pressing decrease the number, and through the digits. Press ENTER to set the number. Remote Interface Operation Use the following command to set the number of samples from your PC 64