Download Evolution 500 User Manual
Transcript
Nicolet Evolution 500 User Manual (1) The amount of energy, or signal, reaching the detector at any wavelength depends on the source energy and the sample absorption at that wavelength. (2) Noise is proportional to the square root of the signal, therefore the amount of noise increases as a proportion of the signal as the level of signal falls; consequently the amount of noise varies across the scan. (3) Where high absorbance coincides with low source energy the noise due to the signal is markedly greater than where high absorbance coincides with high source energy. (4) Noise can be reduced by increasing the time for which a data point is measured (integration time). (5) Noise can be made constant across the scan by varying the integration time at each data point. 9.5.4.1 Standard The Standard scan speed option allows the setting of fixed scan speeds. The maximum rate at which the spectrophotometer can scan is 1200 data points per minute. This equates to 2400 nm/min at 2nm data interval (or 120nm/min at 0.1nm data interval) with an integration time of 15 milliseconds. By slowing the scan speed and/or increasing the data interval the integration time can be increased and consequently noise decreased. 9.5.4.2 Intelliscan Intelliscan is a means of optimising the noise on a spectrum by varying the integration time at each data point and therefore the scan speed over the spectrum. How Intelliscan Works As the energy hits the detector, an analogue signal is produced which is converted to a digital signal by the analogue to digital converter (ADC). For the ADC to work effectively, the signal reaching it has to be within a preset range. If the signal is outside this range then it is amplified. The level of amplification is called the gain. So if a given signal is below the range specified for the ADC, another detector reading is taken and the gain increased. This process continues with the gain increasing until the signal level is within the acceptable range. A measurement is then taken and the spectrophotometer moves on to the next wavelength. The level of gain has no effect on the strength of the signal; it simply multiplies up a weak signal to fit the appropriate ADC range. The level of gain is, however, a direct indicator of the amount of energy hitting the detector which in turn is dependent on the combination of source energy less absorbance of the sample. Therefore the level of gain set at any wavelength can be used to set the integration time required at that wavelength. High energy at the detector requires no gain and therefore a short integration time; low energy requires high gain and therefore a longer integration time. This is what Intelliscan does by intelligently varying the integration time at each data point of the scan depending on the amount of light hitting the detector as measured by the gain required to fill the ADC. 49 Issue 1 (August 2003)