Calibrating for Increased Measurement Accuracy Frequency Response Error Corrections Figure 6-2 Standard Connections for a Response Error Correction for Reflection Measurement 7. To measure the standard when the displayed trace has settled, press SHORT or OPEN . If the calibration kit you selected has a choice between male and female calibration standards, remember to select the sex that applies to the test port and not the standard. The analyzer displays WAIT - MEASURING CAL STANDARD during the standard measurement. The analyzer underlines the softkey that you selected after it finishes the measurement, and computes the error coefficients. NOTE This calibration allows only one standard to be measured. If you press the wrong key for a standard, press RESPONSE again and choose the correct standard. Do not use a thru standard for a reflection response correction. NOTE You can save or store the measurement correction to use for later measurements, that use the same measurement parameters. Refer to the Chapter 4 , “Printing, Plotting, and Saving Measurement Results” for procedures. This completes the response correction for reflection measurements. You can connect and measure your device under test. 6- 13
Related documents
Agilent Technologies 24A User`s guide
Agilent Technologies 8753ES User`s guide
PMM 7010 - MPB Srl
Notice - Agilent Technologies
MANUAL V36-V42
HP 85062A 3.5 mm Electronic Calibration Kit
G-clamp v2 User Manual - University of Pittsburgh
- ftp.nero.com
Specifying Calibration Standards and Kits for Agilent Vector Network
Agilent Technologies 8753ES Specifications
Improving Throughput in Network Analyzer Applications
Improved Throughput in Network Analyzer Applications