PNA calibration kit entry/modification procedure
Use the “Advance Modify Cal Kit Wizard” on the PNA’s
[Calibration] pull down menu to create or modify a calibration kit
for the PNA.
The PNA’s built-in cal kit editor will show a list of all cal kits
files currently available and can be selected for modification. All
current Agilent mechanical calibration kits are preinstalled in the
factory cal kit file directory. If the kit of interest is not on the list,
import the kit (instructions follow). Once on the list, select the
kit, then click on [Edit Kit].
Or, download a PC executable PNA Cal Kit Editor program. The
PC version may not be as updated as the PNA’s built-in version.
Newer versions can process kit data created by an older version.
The reverse may not be true.
Cal Kits can be imported from other directories, the disc drive or
the USB drive. Almost all previous versions of network analyzer
data files can be imported. Some user created 8510 cal kit files
and 87xx cal kit files may not import properly. Check the kit’s
content after download to make sure all data entries are correct.
Note that PNA versions of all the Agilent 8510 calibration kits
were pre-installed in the factory cal kit directory.
With the PC version, select a kit and then select [Edit] >[Modify].
Figure 14. Select and modify cal kit