Download TP01 Manual version 1509
Transcript
Hukseflux Thermal Sensors 1 Theory The following chapter gives a summary of the TP01 theory. A more extensive explanation can be found in an appendix of this manual. As indicated in the introduction, the TP01 design is a modification of the well known non-steady-state probe. This technique utilizes the temperature measurement around a heating wire to analyze soil properties. All non-steady-state probes are based on the same phenomenon: that one can determine the thermal properties of a medium from the temperature response to heating. After an initial transition period, the temperature rise close to the heater depends only on the thermal conductivity of the surrounding medium, and no longer on heat capacity. This method avoids the necessity to reach a true thermal equilibrium with constant temperatures. Non-steady-state techniques are fast and also there is no need for careful sample preparation. Sensors based on this principle are therefore suitable for quick experiments and also for field use. Provided that the thermal mass of the sensor itself is small, during the initial transition period the time response of this type of probe is proportional to the thermal diffusivity of the surrounding medium. TP01 uses a new technique which depends heavily on a very sensitive temperature gradient sensor. A differential temperature sensor (2 thermopiles) measures the radial differential temperature around the central heating wire with recordbreaking sensitivity. This technique is easier to employ than conventional techniques because the interpretation of the signals is very easy. The principle of measurement is clarified in figures 0.1, 1.1 and 1.2. TP01 Manual version 1509 page 8/45