3 SPECIFICATIONS
MELSEC-Q
(c) Processing using moving average
Since the calculation is done for each sampling period, the latest digital
output value can be obtained.
Setting range of moving average is 4 to 60times.
When setting a value out of the setting range, an error (error code 31 )
occurs.
Moving average processing in the case of 4-time setting
Count set for averaging
Temperature
3)
[
Sampling cycle
4)
2)
5)
1)
6)
8)
9)
7)
]
12)
10)
11)
Buffer memory
1st storage
Temperature
conversion value
2nd storage
3rd storage
Time [ms]
Data transition in buffer memory
1st storage
2nd storage
3rd storage
1) + 2) + 3) + 4)
2) + 3) + 4) + 5)
3) + 4) + 5) + 6)
2
2
2
3-7