Download User`s Guide: Chromeleon Operational
Transcript
Dionex IC System OQ/PQ User’s Guide pump configured is an IP20/IP25, IC20/IC25/IC25A, IS25, DX-120, ICS-90, ICS-1000, ICS-1500, ICS-2000 or ICS-3000 Isocratic pump. Eluent Description A Deionized water B 1.0 mM KCl C Deionized water D 1.0 mM KCl 10 mM KCl Stock Solution Preparation: Prepare one liter of 10 mM KCl by weighing 0.746 grams of KCl. Place the KCl in a 1L volumetric flask. Fill flask to the mark with ASTM Type 1 (or better) deionized water (18 MΩ-cm, filtered) water. 1 mM KCl Preparation: Pipette 100 mL of the 10 mM KCl solution into another 1L volumetric flask, filling the flask to the mark with ASTM Type 1 (or better) deionized water (18 MΩ-cm, filtered) water. The 1 mM KCL solution should be freshly prepared daily. Remember to … • Degas all eluents. • Connect eluent lines A and C to one eluent bottle containing 1 L of deionized water equipped with a 3-way mixing tee. • Connect eluent lines B and D to one eluent bottle containing 1 L of 1 mM KCl equipped with a 3way mixing tee. • Pressurize eluent bottles with GP40/IP20/IC20/DX-120 pumps. • Eluent bottles on other pumps do not need to be pressurized. 2.4.2 Preparation of Eluents for use with Absorbance Detection These eluents are for used for qualification with an AD20, AD25 or UVD-170U Absorbance Detector or the UVD-340U or PDA-100 Photodiode Array Detector. Prepare the following eluents and connect them to lines A, B, C and D of the pump. All eluents must be degassed. Eluent bottles must be pressurized when using a GP40, IP20, IC20, DX-120 or ICS-90 pump. Eluent lines A and C can be inserted into one eluent bottle containing 1 L of water. Eluent lines B and D can be inserted into one bottle containing 1 L of 0.3% Acetone in Water. Note: Eluents B, C and D are needed for the Gradient Pump Accuracy Test only and should be prepared only if the pump configured has gradient capabilities. Do not prepare these eluents if the pump configured is an IP20/IP25, IC20/IC25/IC25A, IS25, DX-120, ICS-90, ICS-1000 ICS-1500, ICS-2000 or ICS-3000 Isocratic pump. Eluent Description A Deionized water B 0.3% Acetone in Water C Deionized water 031726-15 OQ PQ Operating Instructions rev 5.11.doc Page 21