Download Whisper User Manual - Motion Lab Systems, Inc.
Transcript
C3D Recordings This section will discuss the gain calibration of sample recordings made with the MA-100 and MA-300 EMG systems using a commercial motion capture system with a proprietary ADC card. For the purposes of this tutorial, the two EMG systems differ in their treatment of the EMG signal as the MA-100 system has a continuously variable signal gain while the MA-300 system offers discrete switched gain levels. This section assumes that the reader is familiar with the stand-alone calibration protocols described in the preceding pages and as a result, descriptions of the individual features of the example EMG systems will not be repeated. Full details of the public C3D file format description together with common parameter information is available from the C3D web site at http://www.c3d.org C3D files are a common biomechanics standard that allows analog and 3D data to be recorded in data files, together with various parameters that describe the data. A standard analog channel consists of a number of analog samples together with two gain factors that are applied to the analog signal to transform the binary samples into real-world values. These gain factors are called GEN_SCALE and SCALE. There is a single GEN_SCALE parameter that is applied to every analog channel – the value of the GEN_SCALE parameter is usually set when the data capture system is installed. Since this scale factor is applied to all analog data it is usually not changed very often – we will need to know its value but will not change it when we calibrate our individual analog channels. The calibration for each analog data channels in a C3D file is controlled by an SCALE factor parameter – this parameter has an individual value for each channel that is directly related to the gain or input range of the channel in combination with the GEN_SCALE value. While the GEN_SCALE and individual SCALE parameters can take any value, it is normal to choose these parameters such that results in the analog channel are scaled in some known value – normally this will default to Volts. The C3D analog parameter UNITS should normally describe the measurement system for each channel. Common values for GEN_SCALE are: • 0.0048828 – the value of a single bit of data from a 12-bit ADC that is measuring a ± 10V input range. An individual channel SCALE value would then be 1.00 to obtain the analog data scaled in Volts. • 0.0024414 – the value of a single bit of data from a 12-bit ADC that is measuring a ± 5V input range. An individual channel SCALE value would then be 1.00 to obtain the analog data scaled in Volts. • 1.000000 – individual channel SCALE values must be set to 0.0048828 or 0.0024414 to obtain analog data scaled in Volts when sampled by a 12-bit ADC that is measuring a ± 10V or ± 5V input range respectively. Many commercial motion capture systems use a standard naming procedure for data recording trials that may not allow you to specify the standard names used by the EMG Testing software. If you are using one of these systems then it may be necessary to copy and rename the C3D files after recording to facilitate the use of the Whisper EMG test software. Variable Gain Systems As described earlier, the calibration procedure is for an MA-100 system is simple. A calibration recording is made using the internal calibration source to apply a known calibration source to each EMG input. Since we know the input signal level (200μV RMS sine wave) all we need to do is directly measure the recorded level of the resulting output signal to calculate a scaling factor to record data scaled in μV. The exact method of generating a C3D file will vary depending on your data collection software however it is important that the analog parameters GEN_SCALE and 30 • A Whisper Tutorial Whisper User Manual