PSI System Controls and Diagnostics A560 loses power No text appears on LCD. A560 User Manual incompatibility the A560 and looped devices necessary; onsult Pyramid Technical Consultants or your supplier. Internal fuse open. Check power LED. Disconnect power supply and any device using 24V output from A560, wait 5 minutes, retry. Locate source of excess current demand and fix. External 24V PSU foldback. Check power LED. Disconnect any device using 24V output from A560, wait 5 minutes, retry. Locate source of excess current demand and fix. Firmware corrupted or Check rear panel absent. LED power up sequence. Contact Pyramid Technical Consultants. You will be advised whether to return the unit or attempt a reprogramming procedure. A560_UM_150413 Page 51 of 67