readings. Data is uploaded to the host in blocks of 100 while the acquisition is in
progress. If you set a buffer of 0, then buffering is disabled and the data is sent to
the host computer as fast as it can be accepted.
Burst count The burst count is the number of readings that are taken after each trigger. The
buffer may be considered as a reserve from which fixed portion of size burst count
is drawn whenever there is a trigger.
10.3.4.3
Other setup
Figure 22. Other setup parameters.
Pulse setup
You can enable the diagnostic pulse output individually for the four channels by
checking any combination of the boxes A to D. This facility is typically used in
scintillation detectors that include a test LED in the housing, to confirm the
complete detector and counting system is working. The pulse width is adjustable
down to 1 nsec and the period down to 1 µsec (corresponding to 1 MHz).
Encoder
You can limit the encoder count to a specific maximum, beyond which it starts
again from zero. This is typical for shaft encoders. Setting zero means there is no
limit to the count.
If there is an index signal available, you can cause this to zero the counter, and
you can choose the polarity of the index.
The counting sense relative to the physical motion sense can be swapped by
checking the Invert Encoder box.
10.3.5 Discriminator sweep
In order to set the detector high voltage and the discriminator levels, you need information about
the distribution of pulse heights. The C400 provides an automatic discriminator sweep facility
that displays a spectrum of pulse heights, allowing you to make informed decisions about the
settings. The discriminator and high voltage settings are reproduced in this display area so that
you can enter the values you choose conveniently. See section 15 for further details.
C400 User Manual
C400_UM_131220
Page 42 of 107