FCC Part 15, Subpart C, Section 15.219 – Unlicensed Low Power Transmitter
Page 23
Procaster, Model: AMTX100
FCC ID: VCJ- AMTX100
EXHIBIT 6. MEASUREMENT UNCERTAINTY
The measurement uncertainties stated were calculated in accordance with the requirements of NIST Technical
Note 1297 and LAB 34.
6.1.
LINE CONDUCTED EMISSION MEASUREMENT UNCERTAINTY
CONTRIBUTION
(Line Conducted)
EMI Receiver specification
LISN coupling specification
Cable and Input Transient Limiter calibration
Mismatch: Receiver VRC Γ1 = 0.03
LISN VRC ΓR = 0.8(9 kHz) 0.2 (30 MHz)
Uncertainty limits 20Log(1+Γ1ΓR)
System repeatability
Repeatability of EUT
Combined standard uncertainty
Expanded uncertainty U
PROBABILITY
DISTRIBUTION
UNCERTAINTY (dB)
9-150 kHz
0.15-30 MHz
Rectangular
Rectangular
Normal (k=2)
+1.5
+1.5
+0.3
+1.5
+1.5
+0.5
U-Shaped
+0.2
+0.3
Std. deviation
-Normal
Normal (k=2)
+0.2
-+1.25
+2.50
+0.05
-+1.30
+2.60
Sample Calculation for Measurement Accuracy in 450 kHz to 30 MHz Band:
uc(y) = √ m∑ui2(y) = + √ (1.52 + 1.52)/3+ (0.5/2) 2 + (0.05/2) 2+0.352 = + 1.30 dB
I=1
U = 2uc(y) = + 2.6 dB
ULTRATECH GROUP OF LABS
3000 Bristol Circle, Oakville, Ontario, Canada L6H 6G4
Tel. #: 905-829-1570, Fax. #: 905-829-8050, Email: [email protected], Website: http://www.ultratech-labs.com
File #: VIDC001_F15C219
September 27, 2007
All test results contained in this engineering test report are traceable to National Institute of Standards and Technology (NIST)