Transcript
PCS-1000 User Manual CONFigure:CURRent Query Description The CONFigure:CURRent query will return the current range unit. Query Syntax CONFigure:CURRent? Return Parameter <string> Returns the current mode and range unit. Query Example CONF:CURR? > “DC 0.01” Note The range that is returned is the base unit. See the table below: Unit Current Range 100 300A 10 30A 1 3A 0.1 300mA 0.01 30mA CONFigure:CURRent[:DC] Set Description This command will set the current mode to DC and set the range. If the range is not specified, then it will not change. Syntax CONFigure:CURRent[:DC] [<Range> | AUTO] Parameter <Range> Current range <NRf>: 0.00000001~305 The unit will automatically be set to the closest range. AUTO Autorange; Only applicable for the ≤3A ranges. Autorange is not supported for the 30A and 300A ranges. 62