Download FX1000
Transcript
CHAPTER 7 CHANNEL ALARM 7.1 ALARM AND CONFIGURATION This instrument has the channel high alarm and low alarm function, and it supports 1 road alarm relay contact output and saves the latest 50 alarm information, including alarm time, cancellaion alarm time, alarm type and alarm channel. Configuration location: configuration - > function configuration - > alarm, and the configuration screen is as follows: Alarm config Chnl Flow H 60000 L 0 Zone 0 Relay Relay 00 00 Exit Channel Choose alarm channel, flow rate, temperature, pressure (optional). Alarm H, alarm L Set parameter value of high alarm and low alarm. Contact Road contact output is optional, and its capacity is 250 VAC/ 3 A, 30 VDC / 3 A (impedance load). The contact type is normally open. Multiple channel alarm can share contact together. Hysteresis Set alarm hysteresis parameter to prevent frequent alarm when signal has oscillation when it is near the alarm value. Alarm description Alarm type High alarm Low alarm Alarm condition Channel value > high threshold channel value Channel value < low threshold channel value 45 Condition for alarm cancellation Channel value< high threshold channel value hysteresis Channel value > low threshold channel value + hysteresis