11 Quality controls Cedex Bio HT Reviewing the status of controls on board 2 Select a rack from the Samples on Rack list and check the positions. “To load a control if rack positions have been predefined” on page B-169 Tip Use Configuration / Lab / Reserve Positions to check the reserved positions. Viewing missing and blocked controls If controls are requested and one or more of the required control(s) are not on board, the missing controls are listed on the Missing & Blocked tab in the Status work area, and also on the Worklist tab in the Orders work area. a To view missing and blocked controls 1 Click Status on the navigation bar. 2 Click the Missing & Blocked tab. A A Click + to open the folder. Controls are listed with their last rack number and position. Alternatively: 1 Click Orders on the navigation bar. 2 Click the Worklist tab. A B A Double-click the control to display information on the status of this control. B The Blocked Status Text gives the reason why the control is blocked. Reviewing a control definition A control definition contains information about a control and the tests for which it is used. a To review a control definition 1 Click Results on the navigation bar. 2 Click the QC tab. 3 Double-click a control name. Operator’s Manual · Version 1.2 B-167